right hand responses enter the system as:
right index finger : b
right middle finger: y
right ring finger : g
right pinky finger : r
Left hand responses enter the system as:
left index finger : e
left middle finger: w
left ring finger : n
left pinky finger : d
The FORP converts each scannertrigger (typically at the start of each TR) to a simulated keyboard "t" output,
The following settings are for the 2x4 response buttons
Make sure the settings on the forp 932 interface match the picture below
2x4 button box (HHSC-2x4-C)
On the forp 932 you should set the correct values for this 8 button device. The model number of the 8 Button Bimanual Curved Lines is "HHSC-2x4-C". Furthermore we need to set the correct mode of operation. We prefer the letter output (BYGREWND) including a "t" for each trigger sent by the MRI scanner (Mode 002 with HID NAR BYRGT). Mode 000 with HID KEY BYRGT works as well, but for compatibility issues with the Matlab Psychtoolbox we prefer the NAR mode by default. Only the response buttons give the possibility to send a trigger on each scanner TR (as a simulated keyboard "t" output), with a trackball or a joystick the simulated keyboard "t" output is not available.
AUTOCONFIGURE
After the buttons boxes are changed (even if the new button box is of the same type) use AUTOCONFIGURE (or Manual config) and re-select the operating mode (in order to make the device adjust the gain properly on all channels (sort of calibration)).
Instructions:
If the AUTOCONFIGURE option doesn't work you may want to try the MANUAL configuration (see below).
Please inform the tech support about the problem. More information about different operation modes:
http://www.curdes.com/technical/932usemodes
The getting started guide:
http://www.curdes.com/technical/932-Getting-Started